Título: Avaliação de algoritmos de cálculo Acuros XB(AXB) e AAA (anisotropic analytical algorithm) para feixes de fótons em planejamentos hipofracionados de mama
Título(s) alternativo(s): Evaluation of calculation algorithms Acuros XB (AXB) and AAA (anisotropic analytical algorithm) for photon bundles in hypofractionated breast plans
Autor: Mirapalheta, Tatiane Escouto 
Primeiro orientador: Peregrino, Antônio Augusto de Freitas
Primeiro membro da banca: Silva, Leonardo Peres da
Segundo membro da banca: Prinzio, Renato Di
Resumo: O objetivo deste estudo foi comparar os algoritmos Acuros XB com o algoritmo AAA, visualizando seu impacto dosimétrico, tanto nos volumes de tratamento quanto no Organ on risk (OAR), especialmente aqueles envolvendo heterogeneidades pulmonares. Diversas estratégias foram utilizadas para avaliar as características do planejamento do tratamento, IC (Índice de Conformidade) e IH (Índice de Homogeneidade), além da distribuição da dose com V95% e V107% para PTV e aumento do PTV, ambos valores regulatórios. aceito em planos de tratamento clínico 3D. Em relação aos órgãos de risco, as avaliações foram feitas observando-se os valores de referência para tratamentos conformacionais e hipofracionados, propostos por QUANTEC e RTOG 1005. Duas metodologias foram utilizadas (UM constante e re-otimização) para avaliações, foram elaborados planos de tratamento baseados em vinte estudos tomográficos de pacientes do sexo feminino com câncer de mama esquerda, campos tangenciais opostos na técnica field in field, da mesma forma energia de fótons 6MV e 10 MV e semi-tangencial em campos de reforço integrados com filtro dinâmico de 45º ou 60º também foram usados. Para UM constante, V107% no PTV mama em que AAA subestima em -7,52% e AAA SCH em -18,71%.Nas avaliações dos OAR, para pulmão esquerdo, entre AAA e o algoritmo Acuros XB a diferença foi 1,62 % para V5Gy, estando AAA supestimando a dose. Enquanto que para AAA SCH a diferença percentual subestima em -18,67 As métricas de avaliação, nesta metodologia, dose média na região cardíaca, AAA apresenta 2,14% e AAA SCH 3,30% em relação a Acuros XB, sobrestimando a dose para essa métrica. Em relação a região delimitada pele, a dose média para AAA se apresenta -5,54% e AAA SCH -11,70%, subestimando a dose em relação a Acuros XB. A maior diferença observada para V107% no PTV mama, na comparação do AAA SCH com o Acuros XB de 34,32%, e para AAA onde o valor encontrado foi de 13,34%, sobrestimando a dose em relação a referência. Para métrica de avaliação clínica, dose média na região cardíaca, AAA apresenta 2,20% e AAA SCH 5,70% em relação a Acuros XB, sobrestimando a dose para essa métrica. Em relação a mama contralateral, para dose máxima, AAA apresenta -8,17% e AAA SCH -12,73% de dose em relação a referência. Em relação a região delimitada pele, a dose média para AAA se apresenta -4,54% e AAA SCH -8,50%, subestimando a dose em relação a Acuros XB. Foram observados valores de UM que de forma bastante relevante, onde encontramos uma diferença de 2,0 UM entre AAA-R e Acuros XB e 6,6 UM para AAA-R SCH e a mesma referência nas tangentes externas, encontramos também uma diferença importante entre as Um dos algoritmos, para os campos de boost, em que no campo boost 2 para AAA-R -2,4 UM e para AAA-R SCH -7,1UM, ambos comparados a Acuros XB. Nas transições clínicas de um algoritmo para outro, estes devem ser acompanhados por estudos e simulações nos sistemas de planejamento, antes de considerar a implementação clínica
Abstract: The objective of this study was to compare the Acuros XB algorithms with the AAA algorithm, visualizing its dosimetric impact in both treatment volumes and Organ On Risk (OAR), especially those involving pulmonary heterogeneities. Several strategies were used to evaluate the characteristics of treatment planning, HF (Conformity Index) and IH (Homogeneity Index), in addition to the dose distribution with V95% and V107% for PTV and PTV increase, both regulatory values. accepted in 3D clinical treatment plans. In relation to the risk organs, the evaluations were made observing the reference values for conformational and hypofractionated treatments, proposed by QUANTEC and RTOG 1005. Two methodologies were used (UM constant and re-optimization) for evaluations, treatment plans were elaborated based on twenty tomographic studies of female patients with left breast cancer, opposing tangential fields in the field in field technique, as well as 6MV and 10MV and semi-tangential photons in integrated reinforcement fields with a 45º or 60º dynamic filter were also used.For a constant, V107% in the PTV breast where AAA underestimates in -7.52% and AAA SC in -18.71%. In the OAR evaluations, between AAA and the Acuros XB algorithm the difference was 1.62% for V5Gy, with AAA being supportive of the dose. While for AAA SCH the percentage difference underestimates by -18.67 The evaluation metrics, in this methodology, mean dose in the cardiac region, AAA presents 2.14% and AAA SCH 3.30% in relation to Acuros XB, overestimating the dose for this metric. Regarding the skin delimited region, the mean dose for AAA presented -5.54% and AAA-SCH-11.70%, underestimating the dose in relation to Acuros XB.The greatest difference was observed for V107%, in the comparison of AAA SCH with Acuros XB of 34.32%, and for AAA where the value found was 13.34%, overestimating the dose in relation to the reference. For clinical evaluation metric, mean dose in the cardiac region, AAA presented 2.20% and AAA SCH 5.70% in relation to Acuros XB, overestimating the dose for this metric. Regarding the contralateral breast, at maximum dose, AAA presented -8.17% and AAA-SCH -12.73% dose relative to reference. Regarding the delimited skin region, the mean dose for AAA presents -4.54% and AAA-SCH -8.50%, underestimating the dose in relation to Acuros XB.It was observed values of UM that in a very relevant way, where we find a difference of 2.0 UM between AAA-R and Acuros XB and 6,6 UM for AAA-R SCH and the same reference in external tangents, we also find an important difference between the one of the algorithms, for the boost fields, in the boost 2 field for AAA-R -2.4 UM and for AAA-R SCH -7,1UM, both compared to Acuros XB.In clinical transitions from one algorithm to another, these should be accompanied by studies and simulations in planning systems, before considering clinical implementation
Palavras-chave: Breast cancer
Hypofractionation
Calculation algorithms
AAA
Acuros XB
Câncer de mama
Hipofracionamento
Algoritmos de cálculo
AAA A
Radioterapia
Algoritmos
Fótons
Mamas - Câncer
Controle de qualidade
Área(s) do CNPq: CNPQ::CIENCIAS DA SAUDE::MEDICINA::RADIOLOGIA MEDICA
Idioma: por
País: BR
Instituição: Universidade do Estado do Rio de Janeiro
Sigla da instituição: UERJ
Departamento: Centro Biomédico::Instituto de Biologia Roberto Alcantara Gomes
Programa: Programa de Pós-Graduação em Física Médica
Citação: MIRAPALHETA, Tatiane Escouto. Avaliação de algoritmos de cálculo Acuros XB(AXB) e AAA (anisotropic analytical algorithm) para feixes de fótons em planejamentos hipofracionados de mama. 2018. 90 f. Dissertação (Mestrado em Física Médica) - Universidade do Estado do Rio de Janeiro, Rio de Janeiro, 2018.
Tipo de acesso: Acesso Aberto
URI: http://www.bdtd.uerj.br/handle/1/12546
Data de defesa: 19-Jun-2018
Aparece nas coleções:Mestrado Profissional em Física Médica

Arquivos associados a este item:
Arquivo TamanhoFormato 
Tatiane Escouto Mirapalheta Dissertacao completa.pdf1,69 MBAdobe PDFBaixar/Abrir Pré-Visualizar


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.